MCA2100R / MCA2100REN Description (January 2005)
Quantum Software Features
- Library-driven nuclide markers with readout in energy or channels
- Log and linear vertical displays with auto-scale
- Library for storing caibration standards information
- ROI markers displayed after peak search is performed
- Qscript-user-defined automation template
ADC
- 3.5 µs successive approximation
- 14-bit resolution
- Discriminators (computer controlled)
- Upper (105% full scale) in increments of 0.41%
- Lower (0% full scale) in increments of 0.41%
- Zero adjustment via computer in increments of 0.41%
- Maximum throughput: 75,000 cps
Spectrum memory
- QCC mode: 256 or 512 channels
- Linear mode: 250, 500, or 1000 channels
- Maximum counts per channel: 224-1 (i.e., 16 million)
Amplifier
- Bipolar, 2 µs (fixed) shaping
- Built-in dead time correction
- No pole-zero adjustment required
- Coarse gain settings: 1×, 2×, 4×, 8×,
- Fine gain settings: 1 to 2.5× in increments of 0.01×
Counting presets
- Real time
- Live time
- Integral of all ROIs
- Integral of selected ROIs
- Gross ROI statistics
- Net ROI statistics
- Total internal amplifier counts (ICR)
- Total SCA counts
Computer control (via Quantum-MCA)
- RS-232 with available baud rates of
2400, 9600, 19200, 38400, 57600, and 115200
- RS-485 with baud rate of 115200
- 10/100 base-T Ethernet with model 2100REN
- Maximum number of units connected to PC: 8 total
Digital stabilizer
- Internal ADC and gain stabilization
Multichannel Scaling
- Input rate: 5 MHz
- Dwell time: 10 ms to 2.1E7
- Dead time: 3 µs between passes and 3 µs between channels
Bias supply
- 0 to +1200 V in 1-V increments (computer controlled)
Preamplifier
- Charge sensitive
- Internal for use with detector PMTs
Battery backup for
- Spectrum
- Setup parameters
- Clock memory
Front panel indicators
- Dead-time meter (3-color LED array)
- Acquire
- Pulse event detect
- Serial communication in progress
- Power on
Rear panel controls and connectors
- Power switch
- 9-pin 12 V DC power connector
- AUX I/O (auxiliary input/output) rear panel connector (female 9-pin D) with the following connections:
+5 VDC output (100 mA maximum output)
Multichannel scaling input (MCS)
Ground
SCA output (TTL)
ADC gate input (TTL signal with software-selectable
positive or negative polarity for coincidence or
anticoincidence mode)
- Fuse housing
- 9-pin D male RS232 connector
- BNC female connector for input to the preamplifier
- SHV female connector for detector bias
- Two RJ-11 connectors for use with RS485
- RJ-45 connector for Ethernet (Model MCA2100R-EN)
Power
- 90-247 VAC, 50-60 Hz with universal power converter (included)
- or 10-18 VDC@1 A for direct input
- Typical power consumption: 15 watts
Weight
Dimensions
- 2.88 inches (7.32 cm) high
- 10.9 inches (27.7 cm) deep
- 7.63 inches (19.4 cm) wide
Software
- 32-bit QuantumMCA compatible with Windows®
Accessories
- 14-pin tube base
- Selection of signal, interface, and detector bias cables
- Sodium iodide detectors
- Quantum NaI, Quantum NaI/D, Quantum Gold for advanced quantitative anlaysis